Gentier 48E real time PCR machine
Model | Gentier48E |
Thermal block | |
Sample capacity(wells) | 48 |
Reaction volume | 5-100ul |
Consumables | 0.2ml 8-strip tubes,0.2ml PCR single tube(Optical flat cap. Clear tube) |
Temperature range | 0℃-100℃ |
Heating/cooling system | Peltier |
Max heating rate | 8.0℃/sec |
Average heating rate | 6.0℃/sec |
Max cooling rate | 5.5℃/sec |
Average cooling rate | 4.0℃/sec |
Temperature accuracy | ±0.1℃ |
Temperature uniformity | ±0.1℃ |
Gradient range | 1℃-40℃ |
Gradient block | 8row |
Special temperature protocol | Gradient PCR, Long PCR, Touch Down PCR |
Heat Lid | |
Temperature range | Room temperature-110℃ |
Optical system | |
Excitation Source | 4LEDs |
Detector | Photodiode |
Detection position | Excitation and scan from lateral |
Detection method | 4 channels scanning at the same time, no edge effect |
Detection time | 2 seconds for 48 wells for all channels |
Range of excitation/emission wavelengths(nm) | 1. 465/510(FAM,SYBR Green I,SYTO9,Eva Green,LC Green) 2. 527/563(HEX,VIC,TET,JOE) 3. 580/616(ROX,Texas Red) 4. 632/664(Cy5) |
Probe Taqman Probe,Molecular beacons probe, scorpion probe |
|
Multiplexing | Up to 4 targets |
Fluorescence Linearity | r≥0.990 |
Fluorescence Dynamic Range | Adjustable |
Performance | |
Sample Linearity | /r/≥0.999 |
Sample repeatability | Ct value CV≤0.5% |
Sample dynamic Range | 1-1010 copies |
Software Functions | |
Data Analysis Modes | Qualitative analysis, Absolute quantification, Relative quantification, Genotyping analysis, Endpoint analysis, Melt curve analysis, High Resolution Melting |
Control Modes | 7touch screen 2. PC direct control 3.WLAN control(one PC can control max 10 unis, and device can be controlled by any PC in the WLAN) |
Data Storage | Upload and download through USB disk, 1000 results can be stored in machine |
Power failure protection | Automaticly start running experiments after power supply, no need wait PC software |
Customize Report | Templetes reserved, report can customized |
Administration Management | Administratior can set functions limits for users |
Fault Management | Fault report and analysis,solution instruction |
LIS connection | CSV,Excel,TXT format data output, open port for LIS connection |
Others | |
PC operating system | Win7, Win 10 |
Communication Port | 1 ethernet and 2 USB |
Foot print(WxDxH) | 260mmX400mmX260mm |
Weight | 11kg |
Power usage | AC100 to 125V/200to 240V(50/60HZ) |
Power consumption | 600VA |
Work environment | Temperature:10℃~30℃ Humidity:20%~85% |
Write your message here and send it to us