Application:The multi-parameter water quality analyzer is an instrument that mainly uses the ion selective electrode measurement method to achieve accurate detection. It can be widely used in power plants, pure water plants, tap water plants, domestic sewage treatment plants, beverage plants, environmental protection departments, industrial water, and aquaculture. , textile industry, wine industry and pharmaceutical industry, epidemic prevention departments, hospitals and other departments of the ion parameter determination.


Product Detail

Product Tags

GENERAL FEATURES

Color High contrast LCD touchscreen, 7.0 inches.

Intelligent operation system provides features including user management, calibration management, method management, data management and etc.

Multi-reading feature allows auto-read, timed-read and continuous-read.

Automatic/Manual temperature compensation ensures accurate results.

Data Storage 1000 sets for each parameter (GLP-compliant).

Data analysis feature helps user review, compare and recalculate results.

Support for USB or RS-232 communication.

Reset feature automatically resumes all settings back to factory default options.

IP54 waterproof.

 

pH

1-6 points calibration with Standard Recognition.

Selectable pH buffer groups, including DIN, NIST, USA, MERK, JIS, GB.

Support for pH standard group customization.

Automatic electrode diagnosis with pH slope and offset display.

 

Ion

1-6 points calibration

Selectable measurement unit, including μg/L, mg/L, g/L, mmol/L, pX, ppm, etc.

Multi-measurement modes are supported, including Direct Reading mode, Standard Addition mode, Sample Addition mode and GRAN mode.

Over 10 methods are built-in, including F-, Cl-, Br-, I-, NO3-, BF4-, NH4+, K+, Na+, Ca2+, Cu2+, Pb2+, Ag+ and etc., user-defined method is supported.

 

Conductivity

1-5 points calibration with Standard Recognition.

Settable parameters, including cell constant, reference temperature (5/10/15/18/20/25°C), temperature compensation coefficient and TDS factor.

Temperature compensation type (none, linear, pure water).

 

PRODUCT SPECIFICATIONS

Model M500T-A
Parameters pH/EC/ISE/Temp. (mV/ORP/pX/Resistivity/TDS/Sal.)
pH Range -2.000 to 20.000 pH
Resolution 0.1, 0.01, 0.001pH
Accuracy ±0.002 pH
Calibration Points Up to 6
Standard Customization Yes
Calibration Reminder Yes
Standard Recognition NIST, GB, DIN, USA, MERK, JIS buffers
Slope Limit Yes
mV/ORP Range -2000.00 to 2000.00 mV
Resolution 0.1, 0.01 mV
Accuracy ±0.1 mV or ±0.03% of

reading whichever is greater

EH ORP Mode Yes
ORP Calibration Points 1 custom point (relative mV)
pX Range -2.000 to 20.000 pX
Resolution 0.1, 0.01, 0.001 pX
Accuracy ±0.002 pX
Calibration Points Up to 6
ISE Range 1E-9 to 9.999E9
Units mol/L, mmol/L, g/L, mg/L, μg/L, ppm
Resolution Up to 4 significant digits
Accuracy ±0.3%
Calibration Points Up to 6
Conductivity Range 0.000 μS/cm to 2000 mS/cm
Resolution 0.001 μS/cm minimum,

various with range selection

Accuracy ±0.5% FS
Reference Temperature 5, 10, 15, 18, 20, 25 ºC
Calibration Points Up to 5
Calibration Reminder Yes
Standard Recognition 10μS/cm, 84μS/cm, 500μS/cm, 1413μS/cm, 12.88mS/cm
Resistivity Range 5.00 Ω·cm~100.00 MΩ·cm
Resolution 0.01 Ω·cm minimum
Accuracy ±0.5% FS
TDS Range 0.000~1000ppt
Resolution 0.001ppm minimum, changed with range
Accuracy ±0.5%
Salinity Type NaCl%
Range (0.00~8.00)%
Resolution 0.01%
Accuracy ±0.1%
Temperature Range -10 to 135 ºC, 14 to 275 ºF
Unit ºC, ºF
Resolution 0.1
Accuracy ±0.1
Measurement Reading Mode Auto Read (Fast, Medium, Slow), Timed, Continuous
Reading Prompts Reading, Stable, Locked
Temp. Compensation ATC, MTC
Data management Data Storage 1000 Groups
GLP Features Yes
Log Management Yes
Inputs pH Electrode BNC(Q9)
DO with Temp. Probe 4-pin aviation connector
Conductivity with Temp. Probe 5-pin aviation connector
Outputs USB USB 2.0 flash memory device, PC
RS 232 printer, scanner, auto-sampler
Display options Backlight Yes
Auto Shutdown 1~60 min, off
IP Rating IP54
Date and Time Yes
General Power AC Adapter, 100-240V AC input, DC24V output
Dimensions 220×195×68 mm
Weight 950g (2.09 lb)

 