How to calibrate an electronic balance

How to calibrate an electronic balance

Electronic balance error can be reduced, but it will not exist. So how to reduce the error of electronic balance?

The error of the electronic balance is getting bigger and bigger. In the case of eliminating the damage of the parts, it is only necessary to calibrate/calibrate the electronic balance.

A considerable number of electronic balance users have not been calibrated for a long time, and they think that the balance can be directly weighed when it shows zero position. (It should be pointed out that when the electronic balance is turned on and displays the zero point, it does not mean that the accuracy of the weighing data of the balance meets the test standard, but only means that the zero position stability of the balance is qualified. Because to measure whether a balance is qualified or not, other technologies need to be considered comprehensively. compliance with the indicator). Due to long storage time, position movement, environmental changes or to obtain accurate measurements, the balance should generally be calibrated before use. There are two types of calibration methods: internal calibration and external calibration.

Internal school (with automatic internal school model):
1. The balance should be preheated, and the time is about 2-3 hours; if the vibration of the balance is relatively large, the sensor reed of the balance may be displaced. It should be preheated for more than 12 hours, so that the balance can automatically repair the spring. slice shift.
2. The balance should be horizontal, otherwise it must be adjusted, and the spirit level shall prevail.
3. When the weighing pan of the balance has no items to be weighed, it should display the zero position stably.
4. Press the “C” key to start the internal calibration function of the balance, the balance will make a small rustling sound, and then the electronic balance will display “C”, indicating that the internal calibration is in progress.
5. When the electronic balance display shows zero position, it means that the electronic balance should have been calibrated.
If there is an error in the calibration, the electronic balance display will display “Err”, the display time is very short, it should be reset, and the above actions should be repeated for calibration.

Outside school:
1. The balance should be preheated, and the time is about 2-3 hours; if the vibration of the balance is relatively large, the sensor reed of the balance may be displaced. It should be preheated for more than 12 hours, so that the balance can automatically repair the spring. slice shift.
2. The balance should be horizontal, otherwise it must be adjusted, and the spirit level shall prevail.
3. When the weighing pan of the balance has no items to be weighed, it should display the zero position stably.
4. Press the “C” key to start the internal calibration function of the balance, and the electronic balance will display “C” later, and put the standard weight of the balance at this time.
5. When the electronic balance display shows the standard weight of the weight, it means that the electronic balance has been calibrated.


Post time: Aug-11-2022