NIR analyzer
Fast : 10-second analysis of all types of samples
High performance : High accuracy,Repeatability,Stability and Reliability
Easy to implement,operate and maintain
Modern technology and a wide range of available calibrations free of charge
User-freiendly software with multiple statistics analysis
Application scope:
Grain,oilseed processing,Flow milling,Starch production,Sugar production,Ethanol production,Raw material of feed,Feed fnished products
Test parameters:
Moisture,Protein,Fat/Oil,Fiber,Ash,Starch,Amino acid,Sugar,Dry matter and more.
Easy operation with user-friendly software:
Versatile results display with warning and action displays for outlier and product limits
Easy and reliable software operation ensures stable analysis results.
User can develop their own calibrations with total autonomy under the guidance of engineer.
Online data analysis and multiple statistics functions.
Hassle-Free maintenance:
The SupNIR 2700 analyzer are designed to be easily maintained by the customer, thereby decreasing down time and maintenance costs.
The lamp has expected >5,000 hours life and can be easily monitored in the program and changed by the user.
The instrument itself is designed for long life and reliability with quality components and no fans, lasers, desiccators or cooling system to fail.
Installation and training:
The SupNIR is designed for easy setup and configuration. Upon delivery, a factory trained product specialist will perform diagnosis to verify instrument performance and then install calibrations and set up products according to customer requirements.
As part of the installation, the NANBEI products specialist will train the operator and managers in all aspects of routine operation, configuration,
diagnostics and data management.
Technical specifications
Wavelength | 1000-1800 nm, SupN IR 2720 |
Light source | Tungsten halogen lamp with expected lifespan more than 5000 hours, User changeable via specific disassemble tool |
M easurement mode | Diffuse reflectance |
Detector | High Performance ultra- cooled InGaAs (Indium Gallium Arsenide) detecor to -25℃, dual stage temperature stabilized |
Optical bandwidth | 10.95 ± 0.3 nm @ 1395.5 nm |
Background noise | Less than 50 μA |
Sample volume | Flexible 50-150 ml, 3 types of sample cup |
Analysis time | 6-10 sec. (30 scans / sample =6s, scan speed= 5 times/ sec.) |
Wavelength accuracy | <0.1 nm to traceable standard reference material |
Ambient temperature | 5-35 ℃ |
Ambient humidity | <90% RH, < 85% RH recommended |
Power supply | 198-242 V, frequency 50 Hz, 0.5 A |
Dimensions | 403 ×391 ×374 mm |
Weight | 18kg;40lbs |